Skip to main content
Article
A Three-Spectrum Method for Accurate Calibration of Bi-plate Zero-order Retarder
Chinese Optics Letters (2014)
  • Zhigang Han, University of Missouri–St. Louis
  • Zhi Xu, University of Missouri–St. Louis
  • Lei Chen, Nanjing University
Abstract
A new spectroscopic method is proposed for the characterization of optical zero-order retarders. It is demonstrated that the retardance as well as the variation of the effective fast axis of a bi-plate zero-order quarter retarder (633 nm) can be obtained with high accuracy in a broadband wavelength range by taking spectra at only three independent angular orientations of the retarder. The calibration results excellently agree with theoretical models, indicating the new method could be used as a simple and reliable way for efficient self-spectral-calibration of optical zero-order retarders.
Disciplines
Publication Date
January 1, 2014
DOI
10.3788/col201412.031202
Citation Information
Zhigang Han, Zhi Xu and Lei Chen. "A Three-Spectrum Method for Accurate Calibration of Bi-plate Zero-order Retarder" Chinese Optics Letters Vol. 12 Iss. 3 (2014) p. 31202 - 31204
Available at: http://works.bepress.com/zhi-xu/4/