Systematic Differences in Electropherogram Peak Heights Reported by Different Versions of the GeneScan Software
Journal of Forensic Sciences
  • Jason R. Gilder, Wright State University - Main Campus
  • Simon Ford
  • Travis E. Doom, Wright State University - Main Campus
  • Michael L. Raymer, Wright State University - Main Campus
  • Dan E. Krane, Wright State University - Main Campus
Document Type
Article
Publication Date
1-1-2004
Find in a Library
Catalog Record
Abstract

DNA profiling using STRs on the 310 and 3100 Genetic Analyzers routinely generates electropherograms that are analyzed with the GeneScan software available from the instrument's manufacturer, Applied Biosystems. Users have been able to choose from three different smoothing options that have been known to result in significant differences in the peak heights that are reported. Improvements in the underlying algorithm of the most recent version of the software also result in significant and somewhat predictable differences in peak height values. Laboratories that have performed validation studies using older versions of GeneScan should either reanalyze the data generated in those validation studies with the newest version of the software or otherwise take into consideration the systematically higher peak height values obtained as they begin following the recommendation of the manufacturer and use the new algorithm.

DOI
10.1520/JFS2003001
Citation Information
Jason R. Gilder, Simon Ford, Travis E. Doom, Michael L. Raymer, et al.. "Systematic Differences in Electropherogram Peak Heights Reported by Different Versions of the GeneScan Software" Journal of Forensic Sciences Vol. 49 Iss. 1 (2004) ISSN: 00221198
Available at: http://works.bepress.com/travis_doom/36/