Proteomics is the study of all proteins in a biological sample. High-pressure liquid chromatography coupled online with mass spectrometry (HPLC/MS) is currently the method of choice for proteomic analysis. Proteins are extracted, separated at the protein or peptide level (after enzymatic digestion), and fractions are analyzed by HPLC/MS. Detection during off-line fractionation is generally conducted using UV−vis, which is not sensitive enough to distinguish fractions having the largest concentration of proteins/peptides and should not be combined prior to HPLC/MS. To overcome this deficiency, we utilize fluorescence or UV-laser induced fluorescence (UV-LIF) detection for measuring proteins/peptides during the off-line fractionation. Fluorescence detection allows low-abundance proteins/peptides that contain aromatic amino acids to be measured. In this study, peptide/protein samples fractionated using ion-exchange chromatography were detected using UV absorbance, fluorescence, and UV-LIF. The results indicated that fluorescence and UV-LIF were able to detect the lower abundance proteins/peptides to give a more representative chromatogram, allowing the analyst to decide which fractions should be combined prior to HPLC/tandem mass spectrometry (MS/MS) analysis.
- Fractionation,
- peptides,
- proteins,
- fluorescence,
- detection,
- absorption