Skip to main content
Article
Design and Application of Robust Rf Pulses for Toroid Cavity NMR Spectroscopy
Journal of Magnetic Resonance
  • Thomas E. Skinner, Wright State University - Main Campus
  • Michael Braun
  • Klaus Woelk
  • Naum I. Gershenzon, Wright State University - Main Campus
  • Steffen J. Glaser
Document Type
Article
Publication Date
4-1-2011
Abstract

We present robust radio frequency (rf) pulses that tolerate a factor of six inhomogeneity in the B-1 field, significantly enhancing the potential of toroid cavity resonators for NMR spectroscopic applications. Both point-to-point (PP) and unitary rotation (UR) pulses were optimized for excitation, inversion, and refocusing using the gradient ascent pulse engineering (GRAPE) algorithm based on optimal control theory. In addition, the optimized parameterization (OP) algorithm applied to the adiabatic BIR-4 UR pulse scheme enabled ultra-short (50 mu s) pulses with acceptable performance compared to standard implementations. OP also discovered a new class of non-adiabatic pulse shapes with improved performance within the BIR-4 framework. However, none of the OP-BIR4 pulses are competitive with the more generally optimized UR pulses. The advantages of the new pulses are demonstrated in simulations and experiments. In particular, the DQF COSY result presented here represents the first implementation of 2D NMR spectroscopy using a toroid probe.

DOI
10.1016/j.jmr.2011.01.026
Citation Information
Thomas E. Skinner, Michael Braun, Klaus Woelk, Naum I. Gershenzon, et al.. "Design and Application of Robust Rf Pulses for Toroid Cavity NMR Spectroscopy" Journal of Magnetic Resonance Vol. 209 Iss. 2 (2011) p. 282 - 290 ISSN: 1090-7807
Available at: http://works.bepress.com/thomas_skinner/3/