Skip to main content
Article
Analysis of a Clinically Failed, Mechanically Intact, Reconstructive Compression Plate
Trauma Cases and Reviews
  • Bharadwaj Cheruvu
  • Sunil Karmacharya
  • Richard T Laughlin, Wright State University
  • Tarun Goswami, Wright State University - Main Campus
Document Type
Article
Publication Date
9-7-2015
Abstract

A reconstructive orthopedic bone plate was submitted for analysis. Traditional failure analysis methods were used to assess the mode of the plate failure. Metallographic investigation of the plate was carried out in this report. Since limited data was available in the literature and clinical data related to subject demography, date of removal as well as reasons for removal unknown, in depth analysis was not possible. However, the plate was received in in-tact condition with minor biological deposits and scratches, it is speculated that the failure of the plate may have been due to biological/clinical reasons, likely infection, rejecting the device due to lack of union at the fracture site.

Comments

Creative Commons License
This work is licensed under a Creative Commons Attribution 4.0 International License.

DOI
10.23937/2469-5777/1510007
Citation Information
Bharadwaj Cheruvu, Sunil Karmacharya, Richard T Laughlin and Tarun Goswami. "Analysis of a Clinically Failed, Mechanically Intact, Reconstructive Compression Plate" Trauma Cases and Reviews Vol. 1 Iss. 2 (2015) ISSN: 2469-5777
Available at: http://works.bepress.com/tarun-goswami/20/