Skip to main content
Article
Biomechanical Comparison of a Proximal Humeral Locking Plate Using Two Methods of Head Fixation
Journal of Shoulder and Elbow Surgery
  • Ross A Schumer, Wright State University
  • Kara Muckley, Wright State University
  • Ronald J Markert, Wright State University
  • Michael J. Prayson, Wright State University
  • Jonathan Heflin
  • Tarun Goswami, Wright State University - Main Campus
Document Type
Article
Publication Date
6-1-2010
Abstract

Hypothesis

Locking plates have emerged as the implant of choice for stabilization of proximal humeral fractures. The biomechanical properties of a locked plating system using smooth pegs vs threaded screws for fixation of the humeral head were compared to test the hypothesis that there would be no biomechanical difference between pegs and threaded screws. Materials and methods

Sixteen pairs of fresh frozen cadaveric humeri were randomized to have a surgical neck gap osteotomy stabilized with a locked plate using threaded screws (n = 8) or smooth pegs (n = 8). The intact contralateral humerus served as a control. Each specimen was tested with simultaneous cyclic axial compression (40 Nm) and torsion (both ±2 Nm and ±5 Nm) for 6000 cycles. All specimens were loaded to failure. Interfragmentary motion and load-displacement curves were analyzed to identify differences between the groups. Our data were then compared to previously published forces across the glenohumeral joint to provide evidence based recommendations for postoperative use of the shoulder. Results

There was a statistically significant difference between test specimens and their paired control (P < .001) in cyclic testing and load to failure. Differences between the smooth pegs and threaded screws were not statistically significant. Discussion

There is no biomechanical difference between locked smooth pegs and locked threaded screws for proximal fragment fixation in an unstable 2-part proximal humeral fracture model. Conclusion

Our study contributes to the published evidence evaluating forces across the glenohumeral joint and suggests that early use of the affected extremity for simple activities of daily living may be safe. Use of the arm for assisted ambulation requiring a crutch, cane, walker, or wheelchair should be determined on a case-by-case basis.

DOI
10.1016/j.jse.2009.11.003
Citation Information
Ross A Schumer, Kara Muckley, Ronald J Markert, Michael J. Prayson, et al.. "Biomechanical Comparison of a Proximal Humeral Locking Plate Using Two Methods of Head Fixation" Journal of Shoulder and Elbow Surgery Vol. 19 Iss. 4 (2010) p. 495 - 501 ISSN: 1058-2746
Available at: http://works.bepress.com/tarun-goswami/1/