Skip to main content
Article
A novel technique of aortomonoiliac AAA repair in patients with a single patent iliac artery: a "stent-graft sandwich"
Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists
  • Suhail Allaqaband, Aurora Sinai/Aurora St. Luke's Medical Centers, Advocate Aurora Health
  • Ashok Kumar
  • Tanvir Bajwa, Aurora Cardiovascular Services, Aurora Sinai/Aurora St. Luke’s Medical Centers, University of Wisconsin School of Medicine and Public Health
Affiliations

St. Luke's Medical Center

Publication Date
10-1-2004
Abstract

PURPOSE: To present a novel technique for endovascular repair of abdominal aortic aneurysms (AAA) in patients with an occluded iliac artery on one side.

TECHNIQUE: In patients with an occluded iliac artery, the main body of a bifurcated modular endoprosthesis is deployed in standard fashion. Another stent-graft of a similar diameter is placed inside the deployed device so that the proximal end of the second endograft is a few millimeters caudal to the leading edge of the first device, which positions the second contralateral gate caudal to the flow divider of the first device. The second stent-graft is also rotated so that its contralateral gate is 180 degrees opposite to the contralateral gate of the first device. This "stent-graft sandwich" effectively mimics an aortomonoiliac device. After deployment of the second device, the attachment sites are dilated with appropriately sized balloons according to standard practice for the particular endoprostheses.

CONCLUSIONS: Endovascular AAA repair utilizing a "stent-graft sandwich" technique is an effective method of treating AAAs when one iliac artery is occluded.

Document Type
Article
PubMed ID
15482028
Citation Information
Allaqaband S, Kumar A, Bajwa T. A novel technique of aortomonoiliac AAA repair in patients with a single patent iliac artery: a "stent-graft sandwich". J Endovasc Ther. 2004 Oct;11(5):550-2.