Simultaneous Normal and Shear Measurements of Nanoconfined Liquids in a Fiber-based Atomic Force Microscope
Review of Scientific Instruments (2008)
  • Peter M. Hoffmann
  • George Matei
  • Steve Jeffery
  • Shivprasad Patil
  • Shah H. Khan
  • Mircea Pantea
  • John B. Pethica
Abstract
We have developed an atomic force microscopy (AFM) technique that can perform simultaneous normal and shear stiffness measurements of nanoconfined liquids with angstrom-range amplitudes. The AFM technique is based on a fiber-interferometric, small-amplitude, off-resonance AFM. This AFM is capable of providing linear quasistatic measurements of the local mechanical properties of confined liquid layers while only minimally disturbing the layers themselves. A detailed analysis of the measurement geometry reveals that shear stiffness measurements are extremely challenging, as even small deviations from perfect orthogonality can lead to data that is very difficult to interpret. We will show ways out of this dilemma and present results that show simultaneous measurement of the shear and normal stiffness of confined liquid layers.
Publication Date
2008
DOI
https://doi.org/10.1063/1.2839913
Citation Information
Peter M. Hoffmann, George Matei, Steve Jeffery, Shivprasad Patil, et al.. "Simultaneous Normal and Shear Measurements of Nanoconfined Liquids in a Fiber-based Atomic Force Microscope" Review of Scientific Instruments (2008)