Abstract

Premature failure of either the nail and/or locking screws with unstable fracture patterns may lead to angulation, shortening, malunion, and IM nail migration. Up to thirty percent of all unreamed nail locking screws can break after initial weight bearing is allowed at 8–10 weeks if union has not occurred. The primary problem this presents is hardware removal during revision surgery. The purposes of our study was to evaluate the relative fatigue resistance of distal locking screws and bolts from representative manufacturers of tibial IM nail systems, and develop a relative risk assessment of screws and materials used. Evaluations included quantitative and qualitative measures of the relative performance of these screws.

Number of Pages
7
Publisher statement
2009 Lanny V. Griffin,Robert M. Harris, and Joseph J. Zubak.
Citation Information
Lanny V Griffin, Robert M Harris and Joseph J Zubak. "Fatigue strength of common tibial intramedullary nail distal locking screws" Journal of Orthopaedic Surgery and Research Vol. 4 Iss. 11 (2009)