Evaluation of the Lumipulse G TP-N Chemiluminescent Immunoassay as a Syphilis Screening Test
Journal of Clinical Microbiology (2017)
  • Daniel A. Ortiz, University of Texas Medical Branch
  • Michael J. Loeffelholz, University of Texas Medical Branch
Abstract
syphilis diagnosis is often aided by the detection of treponemal and nontreponemal antibodies. Automated treponemal antibody detection systems enable high-volume clinical laboratories to perform syphilis screening at a faster pace with lower labor costs. The Lumipulse G TP-N chemiluminescent immunoassay is an automated system that qualitatively detects IgG and IgM antibodies against Treponema pallidum antigens in human serum and plasma. To assess performance characteristics and workflow efficiency, the Lumipulse G TP-N assay was compared to the Bioplex 2200 Syphilis IgG multiplex flow immunoassay. Among the 4,134 routine and HIV samples tested by the two automated assays, the percentage of agreement was excellent at 99.0% (95% confidence interval [CI], 98.6% to 99.2%; κ, 0.89), with the Lumipulse G TP-N having a shorter time to first and subsequent results. All specimens with reactive syphilis screening results were further tested by rapid plasma reagin (RPR) and Treponema pallidum particle agglutination (TP·PA) testing (n = 231). The results from the RPR-reactive samples (n = 82) showed complete concordance with the two automated assays, while the TP·PA assay displayed some discrepancies. The positive percent agreement (PPA) and negative percent agreement (NPA) between the TP·PA test and the Lumipulse G TP-N test were 98.9% and 77.3%, respectively. The Bioplex 2200 Syphilis IgG immunoassay displayed a similar PPA (100%) but a substantially lower NPA (15.9%). Patient chart reviews of discrepant results suggested that the Lumipulse G TP-N assay produced 27 fewer falsely reactive results and can reduce the amount of additional confirmatory RPR and TP·PA testing needed. The analogous performance characteristics of the two automated systems indicate that the Lumipulse G TP-N assay is suitable for high-throughput syphilis screening.
Keywords
  • Lumipulse G TP-N,
  • chemiluminescent enzyme immunoassay,
  • syphilis,
  • treponemal test
Disciplines
Publication Date
November 1, 2017
DOI
10.1128/JCM.00966-17
Citation Information
Ortiz DA, Loeffelholz MJ. Evaluation of the Lumipulse G TP-N Chemiluminescent Immunoassay as a Syphilis Screening Test. J Clin Microbiol. 2017 Nov;55(11):3236-3241. doi: 10.1128/JCM.00966-17. Epub 2017 Sep 6. PMID: 28878003; PMCID: PMC5654907.