Partial-Data Interpolation Method for Arc Handling in a Computed Tomography Scanner
Computerized Medical Imaging and Graphics
  • Jaisingh Rajwade, Philips Healthcare
  • Lester Miller, Philips Healthcare
  • Daniel J. Simon, Cleveland State University
Document Type
Article
Publication Date
7-1-2012
Abstract

X-ray tube arcing in computed tomography scanners causes poor image quality. During the time that the X-ray tube recovers to full voltage after an arc, image data is being collected. Normally this data, acquired at less than full voltage, is discarded and interpolation is performed over the arc duration. We have developed an algorithm that corrects for improper tube voltage, allowing previously discarded data to be used for imaging. The use of voltage corrected data provides improved image quality compared to simple interpolation methods. This improvement is relevant today as the imaging field uses faster scanners with shorter sampling times.

DOI
10.1016/j.compmedimag.2012.04.004
Version
Postprint
Citation Information
Jaisingh Rajwade, Lester Miller, Dan Simon. (2012) Partial-data interpolation method for arc handling in a computed tomography scanner. Computerized Medical Imaging and Graphics, 36(5), 387-395. doi: 10.1016/j.compmedimag.2012.04.004