Skip to main content
Article
Error Analysis of Multi-Needle Langmuir Probe Measurement Technique
Review of Scientific Instruments
  • Aroh Barjatya, Embry-Riddle Aeronautical University
  • William Merritt, Embry-Riddle Aeronautical University
Submitting Campus
Daytona Beach
Department
Physical Sciences
Document Type
Article
Publication/Presentation Date
4-11-2018
Abstract/Description

Multi-needle Langmuir probe is a fairly new instrument technique that has been flown on several recent sounding rockets and is slated to fly on a subset of QB50 CubeSat constellation. This paper takes a fundamental look into the data analysis procedures used for this instrument to derive absolute electron density. Our calculations suggest that while the technique remains promising, the current data analysis procedures could easily result in errors of 50% or more. We present a simple data analysis adjustment that can reduce errors by at least a factor of five in typical operation.

DOI
https://doi.org/10.1063/1.5022820
Publisher
AIP Publishing
Citation Information
Aroh Barjatya and William Merritt. "Error Analysis of Multi-Needle Langmuir Probe Measurement Technique" Review of Scientific Instruments Vol. 89 Iss. 4 (2018) p. 043507 - 043507-4