Skip to main content
Article
Analysis of Laser Fabricated Microjoints Performance in Cerebrospinal Fluid using a Computational Approach
Journal of the Mechanical Behavior of Biomedical Materials
  • Ahsan Mian, Wright State University - Main Campus
  • Jesse Law
  • Golam Newaz
Document Type
Article
Publication Date
1-1-2011
Abstract

Assessment of neural biocompatibility requires that materials be tested with exposure in neural fluids. We have studied the mechanical performance of laser bonded microjoints between titanium foil and polyimide film (TiPI) in artificial cerebrospinal fluid (CSF). The samples were exposed in CSF for two, four and twelve weeks at 37 °C. The laser microbonds showed initial degradation up to four weeks which then stabilized afterwards and retained similar strength until twelve weeks. To understand this bond degradation mechanism better, a finite element modeling approach was adopted. From the finite element results, it was revealed that bond degradation was not due to the hygroscopic expansion of polyimide. Rather, relaxation of the process induced residual stresses may have resulted in weakening of the bond strength as observed from experimental measurements.

DOI
10.1016/j.jmbbm.2010.09.013
Citation Information
Ahsan Mian, Jesse Law and Golam Newaz. "Analysis of Laser Fabricated Microjoints Performance in Cerebrospinal Fluid using a Computational Approach" Journal of the Mechanical Behavior of Biomedical Materials Vol. 4 Iss. 1 (2011) p. 117 - 124 ISSN: 1751-6161