Skip to main content
Article
Accumulation of E. Coli Bacteria in Mini-Channel Flow
Journal of Biomechanical Engineering
  • M. S. Mayeed
  • Ahsan Mian, Wright State University - Main Campus
  • Greg Auner
  • Golam Newaz
Document Type
Article
Publication Date
6-1-2006
Abstract

The objective of this research is to design and optimize a mini/micro-channel based surface-accumulator of E. coli bacteria to be detected by acoustic wave biosensors. A computational approach has been carried out using the state of the art software, CFD -ACE with water as bacteria bearing fluid. E. coli bacteria have been modeled as random discrete particles tracked by solving the Lagrangian equations. The design challenges are to achieve low shear force (pico-N), high concentration at accumulation, and high enough Reynolds number to avoid bacteria swimming. A range of low Reynolds number (Re) has been considered along with the effects of particle boundary interactions, gravity, Saffman lift, etc. More than two orders of magnitude higher concentration at the accumulation than the inlet concentration, and lower shear force of less than pico-N have been achieved in the optimized designs.

DOI
10.1115/1.2187049
Citation Information
M. S. Mayeed, Ahsan Mian, Greg Auner and Golam Newaz. "Accumulation of E. Coli Bacteria in Mini-Channel Flow" Journal of Biomechanical Engineering Vol. 128 Iss. 3 (2006) p. 458 - 461 ISSN: 0148-0731
Available at: http://works.bepress.com/ahsan-mian/2/