Article
A New Method for Measuring the Neutron Lifetime Using an in Situ Neutron Detector
Review of Scientific Instruments
(2017)
Abstract
In this paper, we describe a new method for measuring surviving neutrons in neutron lifetime measurements using bottled ultracold neutrons (UCN), which provides better characterization of systematic uncertainties and enables higher precision than previous measurement techniques. An active detector that can be lowered into the trap has been used to measure the neutron distribution as a function of height and measure the influence of marginally trapped UCN on the neutron lifetime measurement. In addition, measurements have demonstrated phase-space evolution and its effect on the lifetime measurement.
Keywords
- New Method,
- Measuring,
- Neutron Lifetime,
- Situ Neutron Detector
Disciplines
Publication Date
May 30, 2017
DOI
10.1063/1.4983578
Publisher Statement
Author(s) may post the Version of Record (VOR) 12 months after publication. This document was originally published by Review of Scientific Instruments.
Citation Information
Christopher L. Morris, Edith R. Adamek, Leah J. Broussard, Nathan B. Callahan, et al.. "A New Method for Measuring the Neutron Lifetime Using an in Situ Neutron Detector" Review of Scientific Instruments Vol. 88 (2017) ISSN: 0034-6748 Available at: http://works.bepress.com/robert-pattie/8/